FlexiGC
The smart environmental gascromatograph for industrial plant monitoring.
FlexiGC is the carrier-gas-free gas chromatograph for air-quality applications, designed for the continuous and remote monitoring of BTEX (Benzene, Toluene, Ethylbenzene, and Xylene) in compliance with EN14662-3:2015, as well as other analytical compounds. Its architecture combines compactness, accuracy, and ease of use, and it can also include the optional module for hydrogen sulfide (H₂S).
Built with a modular, compact, and versatile design, FlexiGC stands out for its application flexibility, making it suitable for a wide range of operational contexts:
• Monitoring of industrial emissions (perimeter, diffuse, and fugitive)
• Continuous air-quality control in urban or industrial areas
• Monitoring of sensitive locations such as schools, hospitals, and parks
• Support during the remediation of polluted sites
• Support in monitoring odorous emissions
Easy to install in fixed stations as well as in mobile or emergency surveys, FlexiGC does not require laboratories or complex infrastructure and delivers fast, accurate, and reliable analyses anywhere.
Thanks to integrated cloud connectivity and full compatibility with the Pollution Guardian platform, it enables remote control, centralised data management, and the creation of scalable monitoring networks, ensuring maximum operational efficiency and transparency.
| Dimensions | 420 mm × 620 mm × 210 mm |
| Weight | 19 kg |
| Operating temperature | from -40°C to 50°C |
| Analytical control | Analytical module with microcontroller and industrial PC in chassis |
| Data storage | 64 GB internal flash memory |
| Power supply | 100–240 VAC, 50–60 Hz |
| Max power consumption | ~8A (including heaters, PC (2A), check-standard module, and H2S module) |
| Carrier gas | Ambient air, <10 sccm He, H₂, Ar, N₂ available on request with different cabinet configuration |
| Sampling | Sampling flow 50–300 sccm (typ200 sccm) |
| Detector | Ultra-sensitive PID – Photo IonizationDetector (10.6 eV) |
| Analyzed gases | Benzene, Toluene, Ethylbenzene, Xylenes, others on request |
| Analytical range | 0.2–200 ppb (benzene) with 10-min analytical cycle |
| Custom analysis methods | Additional analytical methods available on request |
| Weather station (optional) | Wind speed/direction, barometric pressure, relative humidity, temperature, solar radiation |
| H2S module (Optional) | 0–2.000 ppb (range) |
| Connectivity | Ethernet, Wi-Fi, 4G |
| Instrument control and data access | Local web server accessible via common browsers (IE, Firefox, Chrome) |
Market:
Environment & Safety
Compounds:
(C2H5)C6H5, (CH3)2C6H4, C6H6, C7H8
Technology:
GC, microGC, PID
Brand:
Pollution